Operation
Service Menu
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
3-77
The Flow Calibration submenu is used to calibrate the flow sensor to zero,
span, or restore factory default values. The flow calibration is visible only
when the instrument is in service mode. For more information on the
service mode, see “
Service Mode
” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Flow Calibration
.
FLOW SENSOR CAL:
>ZERO 200
SPAN 1.0000
SET DEFAULTS
RANGE AVG DIAGS ALARM
The Calibrate Flow Zero screen calibrates the flow sensor at zero flow.
Note
The pump must be disconnected before performing the zero
calibration.
▲
●
In the Main Menu, choose Service > Flow Calibration >
Zero
.
Flow Calibration
Calibrate Flow Zero
RESTORE USER DEFAULTS:
RESTORE
RANGE AVG DIAGS ALARM
RESTORE USER DEFAULTS:
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
¨
TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM