Operation
Instrument Controls Menu
3-30
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
time date flags
10:01 06/20/05 FC0088900
10:02 06/20/05 FC0088900
10:03 06/20/05 FC0088900
10:04 06/20/05 FC0088900
©ª
PGUP/DN
§¨
PAN L/R
RANGE AVG DIAGS ALARM
The Erase Log is used to erase all saved data for the selected record type
(not all short records and long records).
●
In the Main Menu, choose Instrument Controls > Datalogging Settings
>
Erase Log
.
The Select Content submenu displays a list of 32 record fields to use and a
submenu list of the analog output signal group choices to choose from.
Choices are Concentrations, Other Measurements, and Analog Inputs (if
the I/O expansion board is installed). This is a temporary list of items for
the selected record type that must be committed via the datalogging menu
before the changes will apply. Note that committing any changes to this list
will erase all currently logged data, as the format of the stored data is
changed.
●
In the Main Menu, choose Instrument Controls > Datalogging Settings
>
Select Content
.
Erase Log
Select Content
ERASE LREC LOG FILE DATA?
ERASE
RANGE AVG DIAGS ALARM
ERASE LREC LOG FILE DATA?
ERASE
ARE YOU SURE YOU WANT TO?
x
PRESS
¨
TO CONFIRM ERASURE