Error Message
Cause
Suggestion
Too few edges
No of edges found in the gated
region is less than the number
of edges required for the
algorithm for calculation.
Increase the record length.
Pos clipping
Vertical scale is not set
properly.
Use Power Autoset to set the vertical scale automatically.
Neg clipping
Vertical scale is not set
properly.
Use Power Autoset to set the vertical scale automatically.
Invalid range Indices
No data in the gated region.
Make sure that the input waveforms are valid.
Empty Input
Input waveform has no data.
Make sure to set all the valid inputs.
Input all DC
Input signal is pure DC.
Change the input signal to AC or add some AC component to
DC signal.
No data in range
No data in the gated region.
Make sure that the input waveforms are valid.
Warning messages displayed on the Power measurement badges
Warning Message
Cause
Suggestion
No data in range
There are no data between the
two cursors.
Place the cursors appropriately.
Warning from Frequency
measurement
Not able to measure frequency
value on the input waveform
due to noisy waveform.
Adjust Hysteresis band to compute edges properly so that
frequency can be measured. The Hysteresis values are set in
the measurement configuration menu Reference Levels panel.
Warning from RMS
measurement
Not able to measure RMS
value on the input waveform
due to noisy waveform.
Adjust Hysteresis band to compute edges properly so that RMS
can be measured. The Hysteresis values are set in the
measurement configuration menu Reference Levels panel.
Warning from MaxElement
measurement
Not able to measure MIN/MAX
value on the input waveform
due to noisy waveform.
Adjust Hysteresis band to compute edges properly so that MIN/
MAX can be measured. The Hysteresis values are set in the
measurement configuration menu Reference Levels panel.
Voltage source expected
Mismatch between input
waveform and measurement.
For example if a current
waveform is used for dv/dt, this
warning is shown.
Make sure a voltage waveform is set as the voltage source
input.
Current source expected
Mismatch between input
waveform and measurement.
For example if a voltage
waveform is used for di/dt, this
warning is shown.
Make sure a current waveform is set as the current source
input.
Warning from RMS
measurement
Not able to measure RMS
value on the input waveform
due to noisy waveform
Adjust the Hysteresis band to compute edges properly so that
RMS can be measured. The Hysteresis values are set in the
measurement configuration menu Reference Levels panel.
References
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help
479
Summary of Contents for 6 series
Page 24: ...Product documents and support 4 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 42: ...Options 22 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 54: ...Install your instrument 34 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 84: ...Getting acquainted with your instrument 64 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 102: ...Configure the instrument 82 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 148: ...Advanced triggering 128 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 154: ...Zooming on waveforms 134 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 438: ...Waveform acquisition concepts 418 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 448: ...Waveform display concepts 428 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 518: ...Index 498 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...