Filter/Limit Results panel (Measurement Settings menu)
Use these settings to apply a High Pass and/or Low Pass filter to block specified frequency band components when taking
measurements. Use the limit controls to set range of measurement values to measure, and the number of measurements to take
(population).
To open the Filter/Limit Results panel:
1. Double-tap a Measurement badge.
2. Tap the Filter/Limit Results panel.
Filter/Limit Results panel fields and controls. Not all items listed in the table may be shown for all measurements; The panel
only shows fields and controls relevant to the selected measurement.
Field or control
Filter Measurement Results Sets whether this measurement's filter and limits settings are Global or Local.
When set to Global (the default), changing anything in this panel causes the same change to all
other measurement filter and limit settings that also have Global selected.
When set to Local, changing anything in this panel only effects this measurement.
High Pass Filter
Blocks the low frequency band and passes only the high frequency band of the waveform.
Select a Butterworth filter order (No Filter (default), 1st, 2nd, or 3rd) and enter the roll-off
frequency in the field.
FP Freq (F1)
High Pass filter cut-off frequency at which the filter magnitude falls by 3 dB.
Low Pass Filter
Blocks the high frequency band and passes only the low frequency band of the waveform.
Select a Butterworth filter order (No Filter (default), 1st, 2nd, or 3rd) and enter the roll-off
frequency in the field.
LP Freq (F2)
Low Pass filter cut-off frequency at which the filter magnitude falls by 3 dB.
Limit Measurement Results Limit taking measurements results to those that are within the specified Min Value and Max
value range.
Limit Measurement
Population
Limit measurements to the specified number of measurements.
Other measurement settings panels.
Measurement Name panel (Measurement configuration menu)
Configure panel (Measurement configuration menu)
on page 165
Reference Levels panel (Measurement configuration menu)
on page 166
Clock Recovery panel (Measurement configuration menu)
Clock Recovery- Advanced Settings configuration menu
on page 172
Gating panel (Measurement configuration menu)
Menus and dialog boxes
174
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help
Summary of Contents for 6 series
Page 24: ...Product documents and support 4 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 42: ...Options 22 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 54: ...Install your instrument 34 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 84: ...Getting acquainted with your instrument 64 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 102: ...Configure the instrument 82 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 148: ...Advanced triggering 128 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 154: ...Zooming on waveforms 134 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 438: ...Waveform acquisition concepts 418 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 448: ...Waveform display concepts 428 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 518: ...Index 498 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...