Turn Off Time. Turn Off Time is the time taken to get the output voltage to low level (close to zero) after the input voltage is
removed.
■
The Maximum Voltage value should be greater than the Input Trigger level.
■
The Maximum Voltage level value sets the waveform vertical settings to approximately one-fourth of the specified level.
■
The Maximum Time field sets the horizontal units so as to display the entire time in which the output is supposed to reach
zero.
■
The Input Trigger level is set based on the 50 % level of the input channel.
■
Turn Off Time sets the trigger slope to rising edge.
Measurement algorithms
476
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help