Turn On Time measurement: Configure panel.
Field or control
Description
Input Source
Selects the channel connected to the input side of the DUT.
Type
Selects the input to output power conversion type. Default is DC-DC.
Label
Sets the name of the measurement. You can use the default name, or double-tap in the field
and change the label using a connected keyboard or the virtual keyboard.
Maximum Voltage
Sets the maximum input voltage, in the range of 1 V to 500 V.
Input Trigger
Sets the input trigger level, in the range of 1 V to 500 V.
Frequency
Sets the DUT input AC signal frequency, in the range of 1 Hz to 1 MHz.
Available when Type = AC-DC
Maximum Time
Sets the maximum time window in which to capture the turn on time measurement (from the
initial state to a steady state), in the range of 1 μs to 500 s, with a resolution of 0.1 μs.
Number of Outputs (1, 2, 3, 4,
5, 6, or 7)
Sets the number of outputs to measure on the DUT (maximum of seven).
Output Source
Selects the channel source to use for the listed output.
Output Voltage
Sets the expected maximum voltage for each listed output, in the range of -6 kV to +6 kV, with a
resolution of 0.001 V. The measurement uses this value to set the vertical scale units for each
output waveform.
The instrument measures the actual maximum voltage on each output, and uses this value to
determine the output On/Off state levels, where On is ≥90% of the measured value, and Off is
≤10% of the measured value.
NOTE. The configuration value for maximum voltage and trigger level will not be remembered when it changes from AC-AC, AC-
DC, DC-AC, and DC-DC modes. It displays the default values.
Menus and dialog boxes
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help
181
Summary of Contents for 6 series
Page 24: ...Product documents and support 4 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 42: ...Options 22 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 54: ...Install your instrument 34 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 84: ...Getting acquainted with your instrument 64 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 102: ...Configure the instrument 82 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 148: ...Advanced triggering 128 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 154: ...Zooming on waveforms 134 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 438: ...Waveform acquisition concepts 418 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 448: ...Waveform display concepts 428 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Page 518: ...Index 498 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...