Flattop2 window
This window has the lowest scallop loss of any of the windows. It also has a wider resolution bandwidth but lower side lobe
attenuation. Also, it is unique because the time domain shape has negative values.
NOTE. The Flattop2 window is useful for high accuracy magnitude measurements for signals that do not require very narrow
bandwidth.
Gaussian window
This is the default window function (see the next figure). It is unique in that the time-domain shape of an exponential Gaussian
function transforms into a Gaussian exponential shape in the frequency domain.
NOTE. This window provides optimal localization in both the time and the frequency domain.