7-8
920011-649
Operator Manual
Troubleshooting
Alarm sounds, and display shows:
FAULT 11
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 11 =
DOOR OPEN
1. If door opens during cycle, call STERIS.*
2. If door is closed, resume cycle by pressing
START
touch
pad. If situation reoccurs, call STERIS.*
Alarm sounds, and display shows:
FAULT 12
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 12 =
SUMP RTD HEATING MALFUNCTION
1. Verify hot tap water (HTW) supply temperature as per
equipment drawing (920-011-656).
2. Try to resume cycle by pressing
START
touch pad. If
situation reoccurs, call STERIS.*
Alarm sounds, and display shows:
FAULT 13
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 13 =
SUMP RTD DEFECTIVE
Try to resume cycle by pressing
START
touch pad. If
situation reoccurs, call STERIS.*
FAULT 14
– Not used.
*Service charges may be incurred. Consult your warranty for details.
FAULT
POSSIBLE CAUSE AND CORRECTION
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...