2
2-1
Installation Verification
Operator Manual
920011-649
2.1 Technical
Specifications
These specifications are intended to describe the technical information
located on identification nameplate (positioned behind processor, on
right side as seen from the front of processor, on electrical supply box
[refer to Figure 3-2]) of the Reliance
™
EPS Processor and to list other
relevant information.
NOTE: On the identification nameplate, the abbreviation "EPS" stands
for Endoscope Processing System.
2.1.1 Permissible
Environmental Conditions
The
Reliance EPS Processor
is designed to give optimal results in
an environment where maximum relative humidity is 80% for
temperatures up to 31°C (88°F), decreasing linearly to 50% relative
humidity at 40°C (104°F).
Pollution degree 2:
Processor must be installed in an environment
where normally only non-conductive pollution occurs but where
occasional, temporary conductivity caused by condensation can
be expected (as determined according to International Standard
EN/IEC 61010-1, Third Edition).
Refer to Reliance dry chemistry package insert for storage
conditions.
Refer to STERIS chemical indicator for Reliance EPS package insert
for storage conditions.
2.1.2 Noise Level
Equivalent Sound Pressure Level at work station (measured 1 m [36"]
away from equipment and at 1.6 m [63"] from ground:
65.4 dB [A]
).
Results were determined according to ISO-3746 1979 Standard:
Acoustics Determination of Sound Power Levels of Noise Sources
Survey Method
.
2.1.3 Voltage, Amperage and
Power Consumption
The Reliance EPS Processor operates either on:
•
208 V, three-phase, 60 Hz, three-wire + PE;
•
230 V, single-phase, 50 Hz, one-wire + n PE;
•
230 V, single-phase, 60 Hz, two-wire + PE;
•
400 V, three-phase, 50 Hz, three-wire + PE;
•
460 V, three-phase, 60 Hz, three-wire + PE.
•
A protective ground conductor is required (Class 1 Equipment).
•
Installation Category is Overvoltage II.
INSTALLATION VERIFICATION
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...