Operator Manual
Processing Instructions
4.14 Shutdown
There is no Shutdown mode on the Reliance EPS Processor. The
processor should remain
ON
at all times except when performing
maintenance or repairs.
If it is necessary to turn
OFF
the processor, wait until cycle is
completed and in "Ready mode", and display shows:
Press bottom portion of
POWER (ON/OFF)
toggle switch to turn
processor
OFF
.
4.15 Power Failure
4.15.1 General
1. If a power failure occurs or if
POWER
(ON/OFF)
toggle switch is
pressed when a cycle is in progress, when power is restored
display shows:
alternating with:
2. Press
ALARM REPLY
touch pad to acknowledge alarm. Display
shows:
3. Press
STOP
touch pad to abort. Processor performs Abort
sequence three times: chamber is filled with filtered tap water
and water is recirculated and then drained (approximately
15 minutes total). Display shows:
alternating with:
READY
MM-DD-YY HH:MM XM
WARNING – ELECTRIC SHOCK
AND/OR
BURN
HAZARD:
Pressing bottom portion of
POWER
(ON/OFF)
toggle
switch to turn off processor
DOES NOT cut off electrical
power.
Lockout/Tagout
building electrical disconnect
switch to turn off power to
processor for maintenance or
repairs.
FAULT 23
CYCLE DELAYED
SEE FAULT CHART
CYCLE DELAYED
STOP TO CANCEL
CANCELLING
CYCLE
PLEASE WAIT
MMM MIN
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...