5-11
Cycle and Control Value Programming
Operator Manual
920011-649
19. Press
ENTER
touch pad to confirm setting. Display shows:
__ blinking
20. Physician ID function allows operator to scan a bar code to
identify physician who used endoscope to be processed.
Physician ID is recorded on cycle printout.
Press
SCROLL arrows
(up or down) to enable (
YES
) or disable
(
NO
) optional Physician ID function.
21. Press
ENTER
touch pad to confirm setting. Display shows:
__ blinking
22. This screen is to set day countdown before next change of
0.2
μm bacterial-retentive filter. Default setting is 90 days. Range
is between 0 and 90 days.
This setting may be changed depending on use frequency of
processor and quality of water. It also must be reset every time
filter is replaced. After a few months of observing alarm pattern,
supervisor may decide to shorten delay as a preventive measure
to be able to change filter
before
FAULT
30 (0.2 μm FILTER
FAILURE) alarm is triggered, or to prevent filling time from
becoming too long (see
, for
details on fault causes and solutions).
To change setting, press
SCROLL arrows
(left or right) to scroll
to numbers and set day countdown.
23. Press
ENTER
touch pad to confirm setting. Display shows:
__ blinking
24. WASHING SELECT screen allows operator to toggle WITH
WASHING/WITHOUT WASHING mode when a SINGLE
endoscope processing cycle or a DOUBLE endoscope
processing cycle is selected.
Press
SCROLL arrows
(up or down) to enable or disable
WASHING SELECT feature (
YES
= Operator
can
elect whether
or not to process a WASHING phase;
NO
= Operator
cannot
elect whether or not to process a WASHING phase).
PHYSICIAN ID
YES
0.2
Μ
M FILTER
IN XX DAYS
WASHING SELECT
YES
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...