Cycle and Control Value Programming
Operator Manual
920011-649
5.1 General
The Reliance
™
EPS Processor is programmed with the
following factory-set operating cycles:
•
SINGLE
endoscope processing cycle – For processing one
endoscope, placed in control handle boot "A", with or without a
WASHING phase.
•
DOUBLE
endoscope processing cycle – For processing two
endoscopes, placed in control handle boots "A" and "B", with or
without a WASHING phase.
•
D-SHORT and D-LONG
– Two decontamination cycles are
programmed for routine maintenance of the processor. Refer to
,
for details.
Reliance
EPS Processing Cycle is composed of a combination of
some, or all, of the following phases:
•
WASHING
– Includes WASH and RINSE segments.
This phase can be omitted by operator if only high level disinfection of
endoscope is required (refer to
) and if
supervisor allowed this selection
(see
). Supervisor can set WASHING phase to be
repeated twice for each
SINGLE
or
DOUBLE
cycles (see
).
•
HIGH LEVEL DISINFECTION (HLD)
•
RINSE
•
AIR PURGE
See
Table 5-1, Cycle Description Chart
,
for details on each cycle.
IMPORTANT: A listing of the Safety Precautions to be observed when operating and servicing this
processor can be found in S
ECTION
1 of this manual. Do not operate the equipment until you have become
familiar with this information.
CYCLE AND CONTROL VALUE PROGRAMMING
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...