Troubleshooting
Alarm sounds, and display shows:
FAULT 35
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 35 =
BOOT B: OVER PRESSURE
This message warns operator that high pressure in control
handle boot "B" could damage endoscope.
IMPORTANT: If high level disinfection solution remains in
sump, call STERIS.*
Possible causes:
1. An endoscope of a diameter larger than template is
processed with insertion tube sleeve(s). Do not use
insertion tube sleeve for these endoscopes (refer to
S
ECTION
).
2. Relief valve is not working properly. Call STERIS.*
3. Relief valve pressure incorrectly adjusted. Call STERIS.*
4.
For a 60 Hz processor
, ensure there is a spacer between
valve cap and valve stem.
5. Try to resume cycle by pressing
START
touch pad. If
situation reoccurs, call STERIS.*
Alarm sounds, and display shows:
FAULT 36
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 36 =
0.2 µm FILTER PRESSURE TOO LOW
Possible causes:
1. Air compressor is not generating adequate pressure. Call
STERIS.*
2. Defective air regulator, call STERIS.*
3. Orange O-ring located in 0.2
μm
bacterial-retentive filter
housing cap is not installed properly. Reinstall O-ring.
4. Orange O-ring located in 0.2
μm
bacterial-retentive filter
housing cap is damaged. Replace O-ring (P117015-006).
5. Verify if hot water inlet check valve is blocked by debris or
damaged. Call STERIS.*
*Service charges may be incurred. Consult your warranty for details.
FAULT
POSSIBLE CAUSE AND CORRECTION
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...