5-3
Cycle and Control Value Programming
Operator Manual
920011-649
5.2 Supervisor Mode
Supervisor mode touch pads are hidden behind plastic control cover
to limit access. Supervisor mode includes following menus:
•
Change Cycle Values
– Includes displays to modify phase
values (see
Cycle Description Chart
, for adjustable
phase values).
IMPORTANT: To help ensure safe and effective operation and
process integrity, some preset cycle values cannot be modified.
•
Miscellaneous Values
– Includes displays to:
- set language selection
- set current time and date
- activate/deactivate bar code reader (accessory), bar code
IDs displays and bar code ID printouts
-
set 0.2 μm
bacterial-retentive filter countdown
- activate/deactivate WASHING select option
- activate/deactivate repeat WASHING phase
- activate/deactivate water sampling
- activate/deactivate low chemical warning
- activate/deactivate automatic D-SHORT
- select automated leak test feature (if optional Automated
Leak Test Assembly is present)
- set enzyme type
- activate/deactivate printout features (duplicate, multiple and/
or extended printouts)
- print values and cycles
•
Prime Pump
– To prime selected cleaner injection pumps
manually.
•
Bar Code Test
(if accessory is present) – To test proper function
of bar code reader.
•
Operator List
– If bar code reader accessory is present, to
program bar code Operator IDs in processor control.
•
Physician List
– If bar code reader accessory is present, to
program bar code Physician IDs in processor control.
5.2.1 Accessing Supervisor
Mode
To access Supervisor mode, proceed as follows (refer to Figure 3-5
for control panel pushbuttons, touch pads and toggle switches):
1. Using a screwdriver, loosen screws and remove control panel cover.
2. Press top portion of
POWER (ON/OFF)
toggle switch to turn
power
ON
. Display shows:
TESTING MEMORY
PLEASE WAIT
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...