6-19
Routine Maintenance
Operator Manual
920011-649
6.12 Rotary Spray Arm
Assembly
See Figure 6-8.
1. Clean processor chamber rotary spray arm assembly as follows:
a. At top of chamber, loosen thumb screw from rotary spray
arm assembly.
b. Remove and lower rotary spray arm assembly.
c. Clean spray arms under running water.
d. Use a fine wire (approximately wire gauge of a paper clip) to
clean out sediment from spray nozzles. Rinse under running
water.
e. Install rotary spray arm assembly into ceiling of chamber.
Tighten thumb screw.
2. To clean rotary spray arm assembly of endoscope processing
support, repeat
Steps b to d
then reinstall.
WARNING
–
ELECTRIC SHOCK
AND/OR BURN HAZARD:
•
Disconnect all utilities to
processor before servicing.
Do not service processor
unless all utilities have
been properly locked out.
Always
follow
local
electrical codes and safety-
related work practices.
•
Pressing bottom portion of
POWER (ON/OFF) toggle
switch to turn off processor
DOES NOT cut off electrical
power.
Lockout/Tagout
building electrical disconnect
switch to turn off power to
processor for maintenance
or repairs.
•
Fasteners and star washers
are used to ensure protective
bonding continuity. Always
reinstall any star washer
which may
have
been
removed during installation
or servicing.
WARNING
–
SLIPPING HAZARD:
To
avoid
slippery
floor
conditions, immediately wipe
up any spilled liquids or
drippage. If spilled liquids or
drippage are detergents or
other chemicals, follow safety
precautions
and
handling
procedures
set
forth
on
detergent or chemical label
and/or Safety Data Sheet
(SDS).
Rotary Spray Arm
Thumb Screw
Figure 6-8. Rotary Spray Arm Assembly
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...