4-22
920011-649
Operator Manual
Processing Instructions
4.4.5 Audible Signal
See Figure 4-20.
When an automated leak test is performed, there may be an audible
signal to notify the operator the control detected a disconnected or a
non-connected hose while a SINGLE or DOUBLE cycle is running. The
audible signal can be either
ignored
or
acknowledged
. Display shows:
and/or:
1. If audible signal is
ignored
, no further action is required. SINGLE
or DOUBLE cycle runs and the following printout informs
operator automated leak test was not performed:
=========================
HH:MM:SS XM
=========================
LEAK TEST 1: NONE
LEAK TEST 2: NONE
=========================
2. If audible signal is
acknowledged
, operator presses
STOP
touch
pad to silence audible signal. Processor drains and display
shows:
Once draining is completed, operator can then choose between
resuming or cancelling cycle. Display shows:
a. If operator chooses to resume cycle without any hoses
connected either to endoscope or processor, filling phase
takes place and automated leak test is attempted a second
time. Go back to
Step 1
. Do not acknowledge audible
signal.
b. If operator chooses to cancel cycle, door can be opened,
leak test hoses connected and a new cycle can be started.
LEAK TEST 1
DISABLED
LEAK TEST 2
DISABLED
DRAINING
PLEASE WAIT
START TO RESUME
STOP TO CANCEL
1
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...