4-18
920011-649
Operator Manual
Processing Instructions
After cycle is completed, disconnect extremity of leak test hose(s)
from endoscope leak test port(s) (see Figure 4-18).
After cycle is completed, when you need to disconnect quick
connect extremity of leak test hose(s) from flexible connection(s),
proceed as follows:
1. With one hand, hold flexible connection (see Figure 4-17).
2. With other hand, push back quick connect locking sleeve and
gently pull out flexible connection. Flexible connection is
released from quick connect (see Figure 4-17).
4.4.2 Pentax Leak Test Adapter
for i10c Series
If reprocessing Pentax i10c Series endoscopes, follow the Pentax
reprocessing instructions for leak test preparation (see Wall Chart
PTPW3028). Use of the Pentax Venting Connector Adapter (OE-C29)
is required for leak testing these endoscopes.
4.4.3 Starting Cycle
The leak test sequence is initiated after the door is closed, the
SINGLE/
DOUBLE
touch pad is pressed and the
START
touch pad is pressed.
This automated leak test occurs during the filling phase and prior to the
recirculation phase.
See
, for details, instructions and
Safety Precautions
on processing cycles.
Figure 4-19.
Pentax Venting Connector
Adapter
(OE-C29)
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...