4-3
Processing Instructions
Operator Manual
920011-649
4.2 General
Processing Guidelines
All personnel involved with the use of this equipment must
carefully review and comply with all the warnings, cautions and
instructions contained in this section.
These instructions are
important to help ensure efficient processing of the flexible
endoscopes, as well as to protect the health and safety of the
personnel using the processor.
The
Reliance Endoscope Processing System Quick-Reference Wall
Chart
(P920011-651) lists the main steps necessary to achieve
successful processing of flexible endoscopes.
Please display the
wall chart near the processor.
IMPORTANT: This wall chart is to be used as a checklist only, and is
in no way meant to replace the thorough reading and understanding
of this manual.
The
Quick Reference Wall Chart - Flow Units
is also available. This
wall chart lists the Flow Units to be used with the Reliance EPS for
specific models of endoscopes.
Please display the wall chart near
the processor.
The
Endoscope Quick Reference Guide
is also available. This guide
contains the list of endoscopes that are compatible with the
processor, as well as details on how to use the list, prepare
endoscopes for processing and attach the Flow Units.
Please refer to
the Endoscope Quick Reference Guide before processing any
endoscopes.
IMPORTANT: The list of devices given in the Endoscope Quick
Reference Guide may not be all-inclusive but reflects the complete
information available to STERIS Corporation at the date of its
publication. Medical devices can be added to the list without further
notice. To obtain the most current list of medical devices that can be
processed in this processor, contact STERIS.
Use the checklist found in
to ensure all steps described in
to
,
are completed.
4.3 Loading
Endoscope Processing
Support
This section contains important information on the proper placement
of endoscopes in the endoscope processing support.
All personnel
involved in the use of this equipment must carefully review and
comply with all the following warnings, cautions and instructions.
IMPORTANT: Ensure all items have been manually cleaned, leak
tested and prepared for immersion before placing them in the
endoscope processing support (refer to the Endoscope Quick
Reference Guide for details).
1. Press top portion of
POWER (ON/OFF)
toggle switch if
processor is not
ON
. Display shows:
WARNING – INEFFECTIVE
PROCESS HAZARD: STERIS
makes no claim of processing
efficacy
when
these
processing instructions are
not followed.
TESTING MEMORY
PLEASE WAIT
WARNING – INEFFECTIVE
PROCESS HAZARD: Failure
to process endoscopes and
accessories within 30 minutes
after manual cleaning may
result
in
ineffective
processing.
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...