4-28
920011-649
Operator Manual
Processing Instructions
�
If optional Automated Leak Test Assembly is present;
�
Appropriate leak test hose is connected to each endoscope
);
�
Quick connect extremity of each leak test hose is connected
to appropriate flexible connection (see
);
�
Pressure is relieved from endoscope(s) (see
);
�
Reliance dry chemistry
container's expiration date is valid,
container is placed in Reliance CDS and the latch is closed on
the Reliance CDS (see
);
�
STERIS chemical indicator
bottle's expiration date is valid and
chemical indicator strip is placed in the clip on control handle
boot "A" (see
).
If all previous items are satisfactory, proceed with
.
4.8 Processing Cycle
There may be up to four cycle options to choose from depending if
WASHING SELECT option is set to YES or NO in Supervisor mode:
•
SINGLE endoscope processing cycle
with washing phase
, to
process one endoscope (one endoscope is placed in control
handle boot "A") through washing and high level disinfection
phases.
•
SINGLE endoscope processing cycle without washing phase
,
to process one endoscope (one endoscope is placed in control
handle boot "A") through a high level disinfection phase only.
•
DOUBLE endoscope processing cycle
with washing phase
,
to process two endoscopes (one endoscope is placed in control
handle boot "A" and one in control handle boot "B") through
washing and high level disinfection phases.
•
DOUBLE endoscope processing cycle
without washing
phase
, to process two endoscopes (one endoscope is placed in
control handle boot "A" and one in control handle boot "B")
through a high level disinfection phase only.
Total nominal cycle time with a washing phase is approximately
42 minutes. Cycle time depends on cycle settings entered in
Supervisor mode. Temperature, pressure and quality of building
water supply can also influence cycle time. Processing cycle
proceeds through cycle phases as described in
Table 5-1
,
Cycle
Description Chart
.
NOTE:
If an error message is displayed at any time during cycle, refer
to
, for possible causes and solutions.
WARNING – INEFFECTIVE
PROCESS HAZARD: Using
washing phase of the cycle
does
not
replace
the
requirement
for
manual
cleaning.
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...