4-15
Processing Instructions
Operator Manual
920011-649
g. Open Reliance CDS.
h. Insert one Reliance dry chemistry container (refer to
Reliance dry chemistry package insert).
i.
Close Reliance CDS and latch.
j.
Insert one STERIS chemical indicator strip in clip on control
handle boot "A" (see
).
k. Ensure pre-processing checklist was completed (see
).
l.
Roll endoscope processing support into chamber. Push
firmly so it connects into manifold connectors at back of
chamber.
m. Manually rotate spray arms on chamber ceiling and under
endoscope processing support to ensure they rotate freely.
n. Close processor door.
o. Press
DOUBLE
touch pad
once
to process a DOUBLE
endoscope processing cycle
with washing
(see
, for more details on how to process cycle).
p. Let cycle proceed through all cycle phases (refer to
Table 5-1,
Cycle Description Chart,
and
for details).
4.4 Optional
Automated Leak Test
See Figures 4-16, 4-17 and 4-18.
IMPORTANT: Ensure leak test is selected in Miscellaneous Values
menu of Supervisor mode (refer to
, for details).
4.4.1 Preparation
Ensure endoscopes and their accessories were prepared and
loaded as described in
Endoscope Quick Reference Guide
and in
. Insert the
Reliance dry chemistry and the STERIS chemical indicator in the
endoscope processing support
AFTER
connection of leak test
hose(s) to flexible connection(s) and
AFTER
connection of leak test
hose(s) to endoscope leak test port(s). Refer to
Endoscope Quick
Reference Guide
,
, and
, for details, instructions and
Safety Precautions
.
IMPORTANT: To ensure all pressure is relieved from endoscope
prior to processing, proceed as follows:
1) Remove endoscope soaking cap from endoscope(s).
2) Reinstall endoscope soaking cap on endoscope(s).
CAUTION – POSSIBLE EQUIPMENT
DAMAGE: Inspect leak test hose
assemblies
and
chamber
flexible
connections
for
presence of water prior to
processing. Water in tubing
during automated leak test may
cause damage to endoscope.
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...