3-7
Component Identification
Operator Manual
920011-649
3.3 Control
IMPORTANT:
Become familiar with all control locations and functions
before operating the Reliance EPS (see Figure 3-5).
The Reliance EPS Processor is equipped with a microprocessor
control system. The control system monitors and automatically
controls all cycle operations and functions. The control system offers
three different modes: Operator, Supervisor and Service mode.
3.3.1 Operator Mode
The Operator mode section of the control panel is used to control the
main functions of the processor: power ON or OFF, cycle selection,
start, stop and abort cycles, and display status or FAULT messages.
NOTE: A description of the control panel symbols is available behind
the control panel cover.
•
POWER (ON/OFF)
toggle switch – press top portion to turn
power
ON
or bottom portion to turn power
OFF
.
•
UNLOCK
toggle switch – press to unlock door.
•
START
touch pad – press once to start a selected cycle.
•
STOP
touch pad – press once to stop/interrupt cycle in
progress.
•
ALARM REPLY
touch pad – press to stop intermittent alarm
buzzer and acknowledge the displayed
FAULT
messages.
•
SINGLE
endoscope processing cycle touch pad – press to
select the preprogrammed SINGLE cycle to process ONE
endoscope. Press once to use WASHING and High Level
Disinfection (HLD) phases; press a second time to process HLD
phase only.
•
DOUBLE
endoscope processing cycle touch pad – press to
select the preprogrammed DOUBLE cycle to process TWO
endoscopes. Press once to use WASHING and HLD phases;
press a second time to process HLD phase only.
NOTE: The choice of selecting the WASHING phase is available
only if WASHING SELECT is set to YES in Supervisor mode. If
WASHING SELECT is set to NO, pressing the touch pad a
second time will have no effect and cycle will always process
WASHING and HLD phases (refer to S
).
•
DECONTAM
cycle touch pad – press repeatedly to toggle
between the standard preprogrammed D-SHORT and D-LONG
cycles.
CAUTION
–
POSSIBLE EQUIPMENT
DAMAGE:
Use a soft damp
cloth and a mild detergent to
clean control panel touch
pads. Splashing water or any
other liquid will damage
control panel touch pads.
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...