4-8
920011-649
Operator Manual
Processing Instructions
11.
If using insertion tube sleeve labeled with a RED dot or a
BLUE dot,
proceed as follows (see
:
a. Insert insertion tube sleeve into control handle boot opening,
as shown in Figure 4-5. Verify insertion tube sleeve handle is
pushed down. Insertion tube must be inserted over sleeve
handle to avoid possible damage.
IMPORTANT: Refer to
Endoscope Quick Reference Guide
to ensure there are no special placement instructions
required for the endoscope being processed.
IMPORTANT: Verify angulation knobs are unlocked and
distal tip is straight before proceeding to next step.
WARNING – INEFFECTIVE
PROCESS HAZARD: When
placing an endoscope in
control handle boot, ensure
endoscope suction barb is
not trapped in the flaps of
universal cord port. Also
ensure light guide cable is
correctly placed between the
flaps of universal cord port.
Figure 4-5. Insertion Tube Sleeves For Small
Diameter Insertion Tubes
Push Down
Sleeve Handle
Flaps
Latch
Insertion Tube
Sleeve With a BLUE
Dot Identification
OR
Insertion Tube
Sleeve With a RED
Dot Identification
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...