1-7
FVS-1000 MMP1
1-10. Installation Procedure
1. Decision of the place to install
Confirm that the following conditions are satisfied.
.
Environmental conditions (Refer to “1-3. Environmental Conditions”)
.
Installation space (Refer to “1-5. Installation Space”)
.
Equipments to supply the power and air (Refer to “1-6. Power Supply and Air Supply Specifications”)
2. Installation of the main unit
(1) Move the unit to the place to install.
(2) Using the handlifter, etc., lift up the unit until the clearance from the caster to the floor becomes 50
mm.
m
.
Use the handlifter of its two claws width is 500 mm or less.
.
Insert the claw to the center position between the casters.
.
Place a rubber sheet, etc. at the part where the claws and the door of the unit contact.
.
When a handlifter is not available, lift up the unit in the following procedure.
1
Place the wooden plates of their thickness are 20 mm under the adjuster feet (four).
2
Turn the adjuster feet with a Phillips screwdriver and lift up the unit.
1-10. Installation Procedure
Claw of the handlifter
Rubber sheet
Claw of the handlifter
Caster
Caster
500 mm or less
Rubber sheet
Adjuster foot
Screwdriver
20 mm
50 mm