![Solartron Mobrey 3595 4C User Manual Download Page 94](http://html1.mh-extra.com/html/solartron-mobrey/3595-4c/3595-4c_user-manual_1314070094.webp)
IMP Commands 3595 4C User Manual Part Two
1-42
35952350 (CH01/AA)
Table 1.18: Calibration responses
e
Code
Meaning
0
Calibration completed.
1
Invalid range.
*
2
EEPROM fault: missing or not working.
3
‘35953Y’ calibration connector not fitted.
*
e = 1 is caused either by an arithmetic error in calculating the calibration coefficients or by the same
coefficients not allowing the full range of the channel to be utilised. Either an incorrect measurement
was made or there is faulty circuitry in the IMP.
+ -
V out
I
out
I
in
CH 1
+ -
V out
I
out
I
in
CH 3
+ -
V out
I
out
I
in
CH 2
+ -
V out
I
out
I
in
CH 4
+ - S
+ - S
35 95 3 Y
Calibra tion
Connec tor
S1
S2
1
2
3
6
5
8
7
4
0
1
2
3
6
5
8
7
4
0
Terminator (for last
IIMP in system)
MULTIMETER
HOST PC
Figure 1.4: Connections to '3595 3Y' calibration block connector
Summary of Contents for 3595 4C
Page 1: ...Precision measurement solutions 3595 4C PC to S Net Interface USER MANUAL 35952350...
Page 2: ...ii 35952350 2001 SOLARTRON MOBREY LIMITED...
Page 6: ...vi 35952350...
Page 7: ......
Page 8: ......
Page 10: ...1 ii 35952350...
Page 12: ...1 iv 35952350...
Page 48: ...Programming the 3595 4C Interface card 3595 4C User Manual Part One 3 20 35952350 CH03 AA...
Page 49: ...3595 4C PC to S Net Interface User Manual PART TWO IMP Commands and Responses...
Page 50: ...2 ii 35952350...
Page 52: ...2 iv 35952350...
Page 108: ...IMP Commands 3595 4C User Manual Part Two 1 56 35952350 CH01 AA...
Page 110: ...Results and Error Formats 3595 4C User Manual Part Two 2 2 35952350 CH02 AA...
Page 122: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 2 35952350 APPX A AA...
Page 126: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 6 35952350 APPX A AA...
Page 127: ......