Function:
Flushes the historical results (FIFO buffer) and returns the FIFO buffer to the ‘free buffer list’.
FB
is a ‘one-shot’ command and has no persistent effect. For further flushing, another
FB
command must be sent.
SD
(
S
ave
D
atabase)
and RD
(
R
estore
D
atabase)
SD
RD
Function:
The SD and RD commands provide for a quick set-up.
The SD command saves the database of a previously configured IMP into a
Flash PROM
.
Later, the RD command can be used to load the current contents of the Flash PROM into the
same database.
When the database is restored, two things happen:
1.
The watchdogs are disabled and
2.
If the channel mode is 80x, the output is set accordingly.
(Refer to Appendix A for details of the database format.)
Compared with the PC saving and restoring the IMP database over S-Net, the above method
is more convenient, but less flexible.
Response:
Stream 3. Single character ‘H’ (
SD
command only). This response confirms that the
database has been written.
Note:
Before storing the database in the Flash PROM, the
SD
command generates an internal HA
(halt) command. This does not change the state of the COntinuous Scan Parameter.
While the database is being restored, there is no communication with the IMP and S-Net.
This causes polling errors, which stop when storage is complete.
It can take up to three seconds for the IMP database to be restored. Therefore after sending
the SD command, any software application should wait three seconds, then clear the error
flags from the 3595 4C Interface card, and then read Stream 3 for an ‘H’ character. Once this
is received, the IMP is ready to resume normal operations.
Summary of Contents for 3595 4C
Page 1: ...Precision measurement solutions 3595 4C PC to S Net Interface USER MANUAL 35952350...
Page 2: ...ii 35952350 2001 SOLARTRON MOBREY LIMITED...
Page 6: ...vi 35952350...
Page 7: ......
Page 8: ......
Page 10: ...1 ii 35952350...
Page 12: ...1 iv 35952350...
Page 48: ...Programming the 3595 4C Interface card 3595 4C User Manual Part One 3 20 35952350 CH03 AA...
Page 49: ...3595 4C PC to S Net Interface User Manual PART TWO IMP Commands and Responses...
Page 50: ...2 ii 35952350...
Page 52: ...2 iv 35952350...
Page 108: ...IMP Commands 3595 4C User Manual Part Two 1 56 35952350 CH01 AA...
Page 110: ...Results and Error Formats 3595 4C User Manual Part Two 2 2 35952350 CH02 AA...
Page 122: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 2 35952350 APPX A AA...
Page 126: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 6 35952350 APPX A AA...
Page 127: ......