IMP Commands 3595 4C User Manual Part Two
1-12
35952350 (CH01/AA)
Table 1.5: Mode Codes for IMP type 2A (Digital)
Code,
m
Mode Set
Comments
000
Skip
700
Digital status
Event count totalise:
740
741
742
-ve going edge.
+ve going edge.
+ve or -ve going edge.
On receipt of a measurement trigger, keeps a
continuous count of events. This count may be
cleared either by the CL command or by setting the
channel mode again to event count totalise.
Event count increment:
750
751
752
-ve going edge.
+ve going edge.
+ve or -ve going edge.
A continuous count of events is kept since the last
trigger. On receipt of the next trigger, the event count
is stopped, the result is sent to the PC, and another
event count is begun.
Event capture:
760
761
762
-ve going edge.
+ve going edge.
+ve or -ve going edge.
Enabled by EV command only.
800
Switch output off (high)
801
Switch output on (low)
For digital output, sets channel to logic ‘1’ or ‘0’.
Frequency Measurement:
900
901
902
903
gate time = 10ms.
gate time = 100ms.
gate time = 1s.
gate time = 10s.
Multiple periods:
910
911
912
913
1 period.
10 periods.
100 periods.
10000 periods.
Returns the time over which periods are counted.
Measurement starts on the negative-going edge.
‘One-shot’ time:
920
921
-ve going start, +ve going stop.
+ve going start, -ve going stop.
Measures -ve pulse width.
Meve pulse width
Note:
-ve edge refers to a negative-going edge, a transition from high to low.
+ve edge refers to a positive-going edge, a transition from low to high.
Event Count
This facility permits a number of events to be counted. The events (transitions) to be
included in the count can be +ve, -ve or both. Two types of count are offered: increment or
totalise. These operate as follows:
Increment
. On receipt of a measurement trigger, an event count is started. Then on
receipt of another measurement trigger, the event count is stopped. The result is sent to
the 3595 4C Interface card and another count begins.
Totalise
. On receipt of a measurement trigger, events are counted until an HA (halt)
command is received or until the channel mode is set to Event Count Increment. In the
latter case, the counter is cleared prior to starting the incremental count. If a CL (clear)
command is received at any time, the event counter is reset to zero. A new count is started
on receipt of the next measurement trigger. The event counter of a particular channel is
also cleared when that channel is set again to the totalise mode.
The maximum number of events that can be recorded per channel by the event-counting
circuitry is 16,777,215, after which the rolls over to zero. No indication of this is given.
Summary of Contents for 3595 4C
Page 1: ...Precision measurement solutions 3595 4C PC to S Net Interface USER MANUAL 35952350...
Page 2: ...ii 35952350 2001 SOLARTRON MOBREY LIMITED...
Page 6: ...vi 35952350...
Page 7: ......
Page 8: ......
Page 10: ...1 ii 35952350...
Page 12: ...1 iv 35952350...
Page 48: ...Programming the 3595 4C Interface card 3595 4C User Manual Part One 3 20 35952350 CH03 AA...
Page 49: ...3595 4C PC to S Net Interface User Manual PART TWO IMP Commands and Responses...
Page 50: ...2 ii 35952350...
Page 52: ...2 iv 35952350...
Page 108: ...IMP Commands 3595 4C User Manual Part Two 1 56 35952350 CH01 AA...
Page 110: ...Results and Error Formats 3595 4C User Manual Part Two 2 2 35952350 CH02 AA...
Page 122: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 2 35952350 APPX A AA...
Page 126: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 6 35952350 APPX A AA...
Page 127: ......