![Solartron Mobrey 3595 4C User Manual Download Page 79](http://html1.mh-extra.com/html/solartron-mobrey/3595-4c/3595-4c_user-manual_1314070079.webp)
3595 4C User Manual Part Two IMP Commands
35952350 (CH01/AA)
1-27
Table 1.12: Connector block codes
Connector Block P/N
Connect Block
Code (
b
)
Connector Block Type
Standard Block
Glanded Block
A
Thermocouple
3595 3A
3596 3A
B
Strain Gauge.
3595 3B
3596 3B
C
Digital.
3595 3C
3596 3C
D
Reed Relay Attenuator.
3595 3D
-
E
Analogue Output.
3595 3E
-
F
Switch.
3595 3F
-
J
Universal.
3595 3J
-
W
Universal Calibration.
3595 3W
-
Y
Analogue Output Calibration.
3595 3Y
-
Z
Calibration.
3595 3Z
-
?
Unknown.
-
-
TR
(TRigger scan)
TR
Notes:
This command does not apply to type 1D IMP devices.
If trigger ‘broadcasting’ is to be used, please read the relevant sub-section in Chapter A of the
3595 4A, 3595 4B, or 3595 9A/9B/9D Operating Manual, before implementation.
Function:
Instructs an ARmed IMP to make a measurement scan. An IMP must already be ARmed to
respond. If CO has already been transmitted, continuous scanning will be started by the TR
command.
On IMP types other than the 1H and 1J, the channels set for event capture, skip, or digital
output, are not affected by this command, but return ‘not measured’ instead. However, the
digital channels (19 and 20) on the 1H and 1J IMP types, when set for digital output, respond
to an ME command by returning the present status.
TRigger will start counting on all channels set to ‘event count increment’ or ‘event count
totalise’, provided counting has not already started. If counting has already started, TRigger
will return to number of events recorded so far; additionally, channels set to ‘event count
increment’ will begin counting from zero.
Response:
Stream 0. Each IMP transmits a 4-byte result for each of its’ channels. If the status-data-
format command instructs the Switch type IMP for compressed data, this data will appear
over stream 3 as a 9-byte quantity.
See also:
AR, EV, SF and (for 1H and 1J IMP types) RM.
Summary of Contents for 3595 4C
Page 1: ...Precision measurement solutions 3595 4C PC to S Net Interface USER MANUAL 35952350...
Page 2: ...ii 35952350 2001 SOLARTRON MOBREY LIMITED...
Page 6: ...vi 35952350...
Page 7: ......
Page 8: ......
Page 10: ...1 ii 35952350...
Page 12: ...1 iv 35952350...
Page 48: ...Programming the 3595 4C Interface card 3595 4C User Manual Part One 3 20 35952350 CH03 AA...
Page 49: ...3595 4C PC to S Net Interface User Manual PART TWO IMP Commands and Responses...
Page 50: ...2 ii 35952350...
Page 52: ...2 iv 35952350...
Page 108: ...IMP Commands 3595 4C User Manual Part Two 1 56 35952350 CH01 AA...
Page 110: ...Results and Error Formats 3595 4C User Manual Part Two 2 2 35952350 CH02 AA...
Page 122: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 2 35952350 APPX A AA...
Page 126: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 6 35952350 APPX A AA...
Page 127: ......