![Solartron Mobrey 3595 4C User Manual Download Page 92](http://html1.mh-extra.com/html/solartron-mobrey/3595-4c/3595-4c_user-manual_1314070092.webp)
IMP Commands 3595 4C User Manual Part Two
1-40
35952350 (CH01/AA)
1.3.6 COMMANDS FOR ANALOGUE OUTPUTS
CH VO
(CHannel Voltage Output)
CH
n
VO
x
Where:
n
is an integer that defines the channel number, in the range 1
≤
n
≤
4
x
is the output voltage in 4-byte floating-point format, in the range -10
≤
x
≤
+10
Note:
This command applies only to IMP type 1D.
Function:
Sets the specified channel to the specified voltage.
Response:
None.
See also:
OS.
CH IO
(CHannel Current)
CH
n
IO
x
Where:
n
is an integer that defines the channel number, in the range 1
≤
n
≤
4
x
is the output current (amps) in 4-byte floating-point format:
•
in the range 0.000
≤
x
≤
0.02, or
•
in the range 0.004
≤
x
≤
0.02, if the split pad on the converter block is made. (See the
IMP Installation Guide)
Note:
This command applies only to IMP type 1D.
Function:
Sets the specified channel to the specified current in amps.
Response:
None.
See also:
OS.
Summary of Contents for 3595 4C
Page 1: ...Precision measurement solutions 3595 4C PC to S Net Interface USER MANUAL 35952350...
Page 2: ...ii 35952350 2001 SOLARTRON MOBREY LIMITED...
Page 6: ...vi 35952350...
Page 7: ......
Page 8: ......
Page 10: ...1 ii 35952350...
Page 12: ...1 iv 35952350...
Page 48: ...Programming the 3595 4C Interface card 3595 4C User Manual Part One 3 20 35952350 CH03 AA...
Page 49: ...3595 4C PC to S Net Interface User Manual PART TWO IMP Commands and Responses...
Page 50: ...2 ii 35952350...
Page 52: ...2 iv 35952350...
Page 108: ...IMP Commands 3595 4C User Manual Part Two 1 56 35952350 CH01 AA...
Page 110: ...Results and Error Formats 3595 4C User Manual Part Two 2 2 35952350 CH02 AA...
Page 122: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 2 35952350 APPX A AA...
Page 126: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 6 35952350 APPX A AA...
Page 127: ......