IMP Commands 3595 4C User Manual Part Two
1-36
35952350 (CH01/AA)
CH TI
(set CHannel TIme-out)
CH
n
TI
p
Where:
n
is an integer that defines the channel number, in the range 1
≤
n
≤
20
p
is an integer that defines the time-out code, in the range 0
≤
p
≤
5
Note:
This command applies only to IMP types 1H, 1J and 2A.
Function:
For a specific channel, sets the time-out period for period measurements. The settings
available are listed in Table 1.17. The default period is 2 seconds (p = 1)
The time-out period is the maximum time any period or ‘one-shot’ measurement is allowed to
take. If the period to be measured exceeds this time, the error ‘period time-out’ is returned
instead of a result. For single period measurement of cyclic signals, the time-out period must
be at least twice that of the measured signal. This ensures both edges of the signal are
within the time-out period.
Response:
None
Table 1.17: Time-out periods
Time-out Code
Time-out Period
0
200ms
1
2s
2
20s
3
50s
4
70s*
5
130s*
*
Available on 1H and 1J IMP types only.
CL
(CLear event totalis counter)
CL
n
Where:
n
is an integer that defines the channel number, in the range 1
≤
n
≤
20
Note:
This command applies only to the digital IMP (type 2A) and the Universal IMP (types 1H and 1J).
Function:
Instructs the IMP to clear its’ event totalise counter and to inhibit counting until the next
ME
or
TR
command, for a specified channel only. The event totalise counter keeps a continuously
updated record of the number of events that have occurred since an initial
ME
or
TR
command. It can only be cleared by a power-down or the
CL
command. Maximum count
value = 16,777,215.
Response:
None.
Summary of Contents for 3595 4C
Page 1: ...Precision measurement solutions 3595 4C PC to S Net Interface USER MANUAL 35952350...
Page 2: ...ii 35952350 2001 SOLARTRON MOBREY LIMITED...
Page 6: ...vi 35952350...
Page 7: ......
Page 8: ......
Page 10: ...1 ii 35952350...
Page 12: ...1 iv 35952350...
Page 48: ...Programming the 3595 4C Interface card 3595 4C User Manual Part One 3 20 35952350 CH03 AA...
Page 49: ...3595 4C PC to S Net Interface User Manual PART TWO IMP Commands and Responses...
Page 50: ...2 ii 35952350...
Page 52: ...2 iv 35952350...
Page 108: ...IMP Commands 3595 4C User Manual Part Two 1 56 35952350 CH01 AA...
Page 110: ...Results and Error Formats 3595 4C User Manual Part Two 2 2 35952350 CH02 AA...
Page 122: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 2 35952350 APPX A AA...
Page 126: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 6 35952350 APPX A AA...
Page 127: ......