![Solartron Mobrey 3595 4C User Manual Download Page 84](http://html1.mh-extra.com/html/solartron-mobrey/3595-4c/3595-4c_user-manual_1314070084.webp)
IMP Commands 3595 4C User Manual Part Two
1-32
35952350 (CH01/AA)
TC
(set
T
hermocouple
C
hecking for open circuit)
TC
n
Where:
n
= 0, thermocouple checking off
n
= 1, thermocouple checking on
Note:
This command applies only to IMP types 1A, 1C, 1E, 1H and 1J.
Function:
With checking enabled, a second measurement follows the thermocouple measurement; this
is to confirm thermocouple integrity. Note that this may slow down the data acquisition rate to
less than 1 per second. When an open circuit is detected, the measurement result is
replaced with the error code. (See Chapter 2.)
Response:
None.
Summary of Contents for 3595 4C
Page 1: ...Precision measurement solutions 3595 4C PC to S Net Interface USER MANUAL 35952350...
Page 2: ...ii 35952350 2001 SOLARTRON MOBREY LIMITED...
Page 6: ...vi 35952350...
Page 7: ......
Page 8: ......
Page 10: ...1 ii 35952350...
Page 12: ...1 iv 35952350...
Page 48: ...Programming the 3595 4C Interface card 3595 4C User Manual Part One 3 20 35952350 CH03 AA...
Page 49: ...3595 4C PC to S Net Interface User Manual PART TWO IMP Commands and Responses...
Page 50: ...2 ii 35952350...
Page 52: ...2 iv 35952350...
Page 108: ...IMP Commands 3595 4C User Manual Part Two 1 56 35952350 CH01 AA...
Page 110: ...Results and Error Formats 3595 4C User Manual Part Two 2 2 35952350 CH02 AA...
Page 122: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 2 35952350 APPX A AA...
Page 126: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 6 35952350 APPX A AA...
Page 127: ......