Results and Error Formats 3595 4C User Manual Part Two
2-12
35952350 (CH02/AA)
2.7 IMP ERROR MESSAGES
Error messages are transmitted by an IMP instead of a 4-byte result. They occur only in response to a
ME
or
TR
command.
An error message is in the form of a 4-byte analogue result with a negative sign and an exponent of 255. In
hexadecimal format, this is represented by any number in excess of ‘FF800000’. In IEEE 754 floating-point
format, this is equivalent to ‘Not a number’.
FF81 xxxx
Analogue overload. The input to an analogue measurement channel has exceeded the
maximum value of the present range.
FF82 xxxx
User thermocouple undefined. Returned when a channel mode is set to 39x or 3Ax, but
the corresponding thermocouple has not been defined. (3595 1H and 1J only)
FF83 xxxx
Out of linearisation range. Returned by a thermocouple channel when the temperature is
out of measurable range.
FF84 xxxx
Ambient temperature range. The IMP has been instructed to use a temperature reference
outside the permitted range (-30
°
C to +80
°
C).
FF85 xxxx
Transducer error. The IMP analogue-to-digital converter is unable to decide a value. This
error is commonly due to an open-circuit input or a large over-voltage.
FF86 xxxx
Open-circuit thermocouple error. Open-circuit thermocouple checking is enabled and the
IMP detected a failure.
FF87 xxxx
Unknown mode, type or range. The IMP has been instructed to select an inapplicable
channel mode or an unsupported integration time.
FF88 xxxx
This error code is unassigned.
FF89 xxxx
Channel number out of range. A channel number, n, outside the permitted range has
been specified. The permitted range is 1
≤
n
≤
10 (for IMP type 3595 1B) or 1
≤
n
≤
20 (for
IMP types 3595 1A, 1C, 1H, 1J and 2A).
FF8A xxxx
System zero error. The IMP is unable to perform a drift correction.
FF8B xxxx
System calibration corrupt. The calibration constants (including backup values) held in
non-volatile memory show 'read errors’. Non-volatile memory is either corrupted or not
fitted.
FF8C xxxx
Strain gauge not initialised. The IMP database must be loaded with the gauge offset
resistance and factor. See the ‘Command Directory’ entry for
IN
. (Relevant only to IMP
type 3595 1B.)
FF8D xxxx
Digital result pending. (Relevant only to IMP types 3595 1H and 1J.)
FF8E xxxx
Period time-out. The IMP has tried to measure a time period but took longer than the
maximum time allowed (as set by the
CH TI
command).
FFFF xxxx
Not measured. The channel has been set to ‘skip’, ‘digital output’ or ‘event capture’.
The remainder of possible error codes are unassigned.
Summary of Contents for 3595 4C
Page 1: ...Precision measurement solutions 3595 4C PC to S Net Interface USER MANUAL 35952350...
Page 2: ...ii 35952350 2001 SOLARTRON MOBREY LIMITED...
Page 6: ...vi 35952350...
Page 7: ......
Page 8: ......
Page 10: ...1 ii 35952350...
Page 12: ...1 iv 35952350...
Page 48: ...Programming the 3595 4C Interface card 3595 4C User Manual Part One 3 20 35952350 CH03 AA...
Page 49: ...3595 4C PC to S Net Interface User Manual PART TWO IMP Commands and Responses...
Page 50: ...2 ii 35952350...
Page 52: ...2 iv 35952350...
Page 108: ...IMP Commands 3595 4C User Manual Part Two 1 56 35952350 CH01 AA...
Page 110: ...Results and Error Formats 3595 4C User Manual Part Two 2 2 35952350 CH02 AA...
Page 122: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 2 35952350 APPX A AA...
Page 126: ...Database Structure 3595 1H and 1J IMP 3595 4C User Manual Part Two A 6 35952350 APPX A AA...
Page 127: ......