23. CHECKS AND ADJUSTMENTS
148
5. Use manual sighting to accurately sight the
target.
"11.3 Manually Sighting the Target"
6. Press
[OK]
.
Press
[STOP]
to stop the measurement.
7. Offset value (H, V) is obtained from the set offset
value (H, V) and the measurement results. The
offset value is a constant value that indicates the
number of degrees of misalignment between the
center of the telescope reticle and the center of
the CCD sensor. If the offset value obtained from
the measurement result is significantly larger than
the set offset value, press
[ESC]
and resight the
target.
If the offset value (H, V) obtained from the
measurement results continues to be significantly
large after repeated checks, adjustment is
necessary. Go to step 8.
If one of the offset values exceeds the range, an
error message appears on the screen. Contact
your Sokkia agent to perform the adjustment.
8. Press
[OK]
to renew the offset value.
• Press
[INIT]
to return to the default settings.