Neck
Neck
kV
130
Effective mAs
85
Slice collimation
2.0 mm
Slice width
5.0 mm
Feed/Rotation
18.0 mm
Rotation time
0.8 sec.
Kernel
B50s
Increment
5.0 mm
Image order
cr-ca
CTDI
Vol
9.7 mGy
Patient positioning is very important for artifact-free
images. The thoracic girdle should be positioned as far
as possible in the caudal direction. This can be done
using a strap with a permanent loop or Velcro fastener
at its end. The ends of the strap must be attached to
the patients wrists. Then the strap must be wrapped
around the patients feet with his legs extended and
under tension. The entire thoracic girdle is thus pulled
toward the patients feet.
Contrast medium IV injection
Start delay
45 sec.
Flow rate
3.0 ml/sec.
Total amount
100 ml
99