39
Children
Tips
• Children, who are more than 6 years old, should be
scanned with an adult protocol.
• The second recon job is defined with kernel C60s
and with an overlap for visualizing bone structures
with MPR.
• If the study is focused on bone structure only, the
mAs can be reduced to 20 for all age groups.
Contrast medium IV injection
Start delay
exam dependent
Flow rate
dependent upon needle size/Access site
Total amount
1 – 2 ml per kg of body weight
Age
kV
mAs
CTDI
Vol
(mGy)
< 3 years
110
356.6
3 years – 6 years
110
50
9.5
MPRthick: Sinus
coronal
Image thickness
2
Distance between images
2
Number of images
30
For the 2
nd
reconstruction the Autoload into the MPR
Range on the 3D Card is activated. The images will
be automatically loaded into 3D, MPR, and a coronal
MPRthick Range will pop up. Please notice, if you are
not satisfied with the Range preset adapt the para-
meters to your needs and link them to the series.