
143
Abdomen
AbdRoutine
kV
130
Effective mAs
95
Slice collimation
2.0 mm
Slice width
5.0 mm
Feed/Rotation
15.0 mm
Rotation time
0.6 sec.
Kernel
B41s
Increment
5.0 mm
Image order
cr-ca
CTDI
Vol
10.8 mGy
You could repeat the same protocol simply by clicking
the chronicle with the right mouse button for “repeat“.
E.g. when both non-contrast and contrast studies are
required.
Contrast medium IV injection
Start delay
50 – 60 sec.
Flow rate
4.0 ml/sec.
Total amount
100 ml
Summary of Contents for SOMATOM Emotion 6
Page 1: ...SOMATOM Emotion 6 Application Guide Routine Protocols Software Version A70 ...
Page 23: ...23 General ...
Page 43: ...43 Children ...
Page 59: ...59 Children ...
Page 89: ...Head 89 ...
Page 102: ...102 Shoulder Overview Shoulder Spiral mode for HiRes bone studies and soft tissues ...
Page 139: ...139 Abdomen ...
Page 167: ...167 Pelvis ...
Page 175: ...175 Spine ...
Page 185: ...185 Extremities ...