58
ThoraxSeqHR
Indications:
Sequence mode for High Resolution lung studies, e.g.
interstitial changes of the lung parenchyma, etc.
ThorHR
kV
110
mAs
*
Slice collimation
1.0 mm
Slice width
1.0 mm
Feed/Scan
10.0 mm
Rotation time
1.0 sec.
Kernel
B90s
Image order
cr-ca
Tips
• Children with a body weight of more than 54 kg
should be examined with an adult protocol.
• If you want to acquire the patient at full inspiration
or full expiration, you should practice the breathing
with the patient a few times before beginning the
scan, so that reproducibility may be improved.
* The mAs should be adjusted to the body weight
of the child.
Body weight
kV
mAs
CTDI
Vol
(mGy)
< 35 kg
110
36
0.3
35 – 54 kg
110
120
0.9
Children