Head
In this protocol, two recon tasks are predefined for
image reconstruction for both soft issue and bone struc-
ture. If one of them is not required, you can delete it
by clicking the icon of the recon task on the chronicle
with the right mouse button and select “delete recon
job“. You should do this before you start the recon-
struction.
When performing Orbit studies, the patient should
be told to keep his/her eyes closed. This will reduce the
possibility of small streak artifacts from eye movement.
If the study is focused on bone structure only, you can
reduce the mAs to 20.
Contrast medium IV injection
Start delay
45 sec.
Flow rate
2.0 ml/sec.
Total amount
75 ml
91
MPRthick: Sinus
coronal
Image thickness
2
Distance between images
2
Number of images
30
For the 2
nd
reconstruction the Autoload into the MPR
Range on the 3D Card is activated. The images will
be automatically loaded into 3D, MPR, and a coronal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link
them to the series.