141
Abdomen
AbdRoutine
kV
130
Effective mAs
95
Slice collimation
2.0 mm
Slice width
5.0 mm
Feed/Rotation
15 mm
Rotation time
0.8 sec.
Kernel
B41s
Increment
5.0 mm
Image order
cr-ca
CTDI
Vol
10.8 mGy
You could repeat the same protocol simply by clicking
the chronicle with the right mouse button for “repeat“.
E.g. when both non-contrast and contrast studies are
required.
Contrast medium IV injection
Start delay
50 – 60 sec.
Flow rate
4.0 ml/sec.
Total amount
100 ml