LungLowDose
kV
110
Effective mAs
16
Slice collimation
2.0 mm
Slice width
5.0 mm
Feed/Rotation
12.0 mm
Rotation time
0.8 sec.
Kernel
B50s
Increment
5.0 mm
Image order
cr-ca
CTDI
Vol
1.3 mGy
111
You could repeat the same protocol simply by clicking
the chronicle with the right mouse button for “repeat“.
E.g. when both non-contrast and contrast studies are
required.
For lung cancer staging, this protocol can be combined
with protocol “Neck Routine“.
Contrast medium IV injection
Start delay
30 sec.
Flow rate
2.5 ml/sec.
Total amount
50 – 70 ml
Thorax