AngioRoutine
2
nd
reconstr.
kV
110
Effective mAs
90
Slice collimation
1.0 mm
Slice width
5.0 mm
1.25 mm
Feed/Rotation
9.0 mm
Rotation time
0.8 sec.
Kernel
B31s
B31s
Increment
5.0 mm
0.8 mm
Image order
cr-ca
CTDI
Vol
8.2 mGy
Do not administer oral contrast medium, as this
impairs the editing of MIP/SSD/VRT images. Use water
instead if necessary.
* CARE Bolus may be used to optimize the bolus timing.
Set the ROI for monitoring scan in the abdominal
aorta at the beginning of the scan range with trigger-
ing threshold of 120 HU, or use manual triggering.
MIPthin: AngioRoutine
sagittal
Image thickness
3
Distance between images
3
Number of images
60
For the 2
nd
reconstruction the Autoload into the
MIPthin Range on the 3D Card is activated. The images
will be automatically loaded into 3D, MIPthin, and
sagittal MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
Contrast medium IV injection
Start delay
20 – 25 sec.*
Flow rate
3.0 – 3.5 ml/sec.
Total amount
100 – 120 ml
161
Abdomen
Summary of Contents for SOMATOM Emotion 6
Page 1: ...SOMATOM Emotion 6 Application Guide Routine Protocols Software Version A70 ...
Page 23: ...23 General ...
Page 43: ...43 Children ...
Page 59: ...59 Children ...
Page 89: ...Head 89 ...
Page 102: ...102 Shoulder Overview Shoulder Spiral mode for HiRes bone studies and soft tissues ...
Page 139: ...139 Abdomen ...
Page 167: ...167 Pelvis ...
Page 175: ...175 Spine ...
Page 185: ...185 Extremities ...