74
ExtrHR
Indications:
Spiral mode for High Resolution bone studies, e.g.
tumors, posttraumatic changes, orthopedic indications,
etc.
Two recon jobs are predefined for image reconstruc-
tion of the soft tissue and bone structure.
A range of 10 cm will be covered in 28.3 sec.
ExtrHR
2
nd
reconstr.
kV
110
Effective mAs
*
Slice collimation
0.5 mm
Slice width
3.0 mm
0.63 mm
Feed/Rotation
3.0 mm
Rotation time
0.8 sec.
Kernel
B41s
U90s
Increment
3.0 mm
0.3 mm
Image order
cr-ca
* The mAs should be adjusted to the body weight
of the child.
Body weight
kV
mAs
CTDI
Vol
(mGy)
< 15 kg
110
35
3.3
15 – 34 kg
110
50
4.7
35 – 54 kg
110
80
7.5
Children
For the 2
nd
reconstruction the Autoload into the MPR
Range on the 3D Card is activated. The images will
be automatically loaded into 3D, MPR, and a coronal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link
them to the series.