121
Thorax
ThoraxCombi
2
nd
reconstr.
kV
130
Effective mAs
70
Slice collimation
1.0 mm
Slice width
5.0 mm
1.25 mm
Feed/Rotation
9.0 mm
Rotation time
0.6 sec.
Kernel
B41s
B70s
Increment
5.0 mm
0.8 mm
Image order
cr-ca
CTDI
Vol
9.2 mGy
If you want to reconstruct thin slices in every 2 mm
or 10 mm instead of 0.8 mm as predefined, simply
change the increment before image reconstruction.
Contrast medium IV injection
Start delay
25 sec.
Flow rate
2.5 ml/sec.
Total amount
80 ml
MPRthin: ThoraxCombi
coronal
Image thickness
3
Distance between images
3
Number of images
60
For the 2
nd
reconstruction the Autoload into MPRthin
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MPRthin, and a coronal
MPRthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.