42
NeckRoutine
Indications:
Spiral mode for routine neck studies, e.g. tumors,
lymphoma, abscesses, etc.
A range of 15 cm will be covered in 8.3 sec.
Neck
kV
110
Effective mAs
*
Slice collimation
2.0 mm
Slice width
3.0 mm
Feed/Rotation
18.0 mm
Rotation time
0.8 sec.
Kernel
B50s
Increment
3.0 mm
Image order
ca-cr
Tips
• Children, who are more than 6 years old should be
scanned with an adult protocol.
Contrast medium IV injection
Start delay
exam dependent
Flow rate
dependent upon needle size/Access site
Total amount
1 – 2 ml per kg of body weight
* The mAs should be adjusted to the age of the child.
Age
kV
mAs
CTDI
Vol
(mGy)
< 3 years
110
352.7
3 years – 6 years
110
50
3.9