51
Thorax
2
nd
reconstr.
kV
110
Effective mAs
*
Slice collimation
2.0 mm
Slice width
5.0 mm
5.0 mm
Feed/Rotation
18.0 mm
Rotation time
0.8 sec.
Kernel
B41s
B60s
Increment
5.0 mm
5.0 mm
Image order
cr-ca
Tips
• Children with a body weight of more than 55 kg
should be examined with an adult protocol.
• The first and second recon jobs are defined for
visualization of the mediastinum and the lungs,
respectively.
Contrast medium IV injection
Start delay
exam dependent
Flow rate
dependent upon needle size/Access site
Total amount
1 – 2 ml per kg of body weight
* The mAs should be adjusted to the body weight
of the child.
Body weight
kV
mAs
CTDI
Vol
(mGy)
< 15 kg
110
13
1.0
15 – 24 kg
110
17
1.3
25 – 34 kg
110
26
2.0
35 – 44 kg
110
43
3.3
45 – 55 kg
110
60
4.7
Children