177
Spine
SpineObeseSeq
Indications:
Sequence mode for spine studies of the obese
patients when higher dose are needed, e.g. prolapse,
degenerative changes, trauma, tumors etc.
This protocol contains three ranges:
L3-L4, L4-L5, L5-S1.
The scan length of each range is 2.8 cm
Three different typical gantry tilts are predefined:
for L3-L4: 0°, for L4-L5: +5° and for L5-S1: +15°
SpineObeseSeq
L3-4
L4-5
L5-S1
kV
130
mAs
320
320
370
Slice collimation
1.0 mm
Slice width
2.0 mm
Feed/Scan
6.0 mm
Rotation time
1.0 sec.
Kernel
B31s
Cycle time
2.5 sec.
Image order
cr-ca
CTDI
Vol
42.2 mGy 42.2 mGy 48.8 mGy
Summary of Contents for SOMATOM Emotion 6
Page 1: ...SOMATOM Emotion 6 Application Guide Routine Protocols Software Version A70 ...
Page 23: ...23 General ...
Page 43: ...43 Children ...
Page 59: ...59 Children ...
Page 89: ...Head 89 ...
Page 102: ...102 Shoulder Overview Shoulder Spiral mode for HiRes bone studies and soft tissues ...
Page 139: ...139 Abdomen ...
Page 167: ...167 Pelvis ...
Page 175: ...175 Spine ...
Page 185: ...185 Extremities ...