131
Thorax
ThoraxHR
2
nd
reconstr.
kV
130
Effective mAs
100
Slice collimation
1.0 mm
Slice width
5.0 mm
1.0 mm
Feed/Rotation
9.0 mm
Rotation time
0.8 sec.
Kernel
B41s
B70s
Increment
5.0 mm
0.7 mm
Image order
cr-ca
CTDI
Vol
13.2 mGy
With studies of interstitial changes in the lungs, contrast
medium is not necessary.
MPRthick: ThoraxHR
coronal
Image thickness
2
Distance between images
2
Number of images
70
For the 2
nd
reconstruction the Autoload into the MPR
Range on the 3D Card is activated. The images will
be automatically loaded into 3D, MPR, and a coronal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.