56
ThoraxCombi06s
Indications:
Combining thin slice lung and routine thorax studies
with one spiral scan. E. g. thorax studies in general
and interstitial changes in the lungs. There are two
recon jobs predefined. The first one, with 3 mm slice
thickness, for the mediastinum, the second recon-
struction, with 1.25 mm slice thickness, for the lung,
using a 0.6 sec. rotation time.
A range of 15 cm will be covered in 11.2 sec.