159
AngioFast
kV
110
Effective mAs
90
Slice collimation
2.0 mm
Slice width
3.0 mm
Feed/Rotation
21.0 mm
Rotation time
0.6 sec.
Kernel
B31s
Increment
3.0 mm
Image order
cr-ca
CTDI
Vol
7.0 mGy
Do not administer oral contrast medium, as this
impairs the editing of MIP/SSD/VRT images.
* CARE Bolus may be used to optimize the bolus timing.
Set the ROI for monitoring scan in the abdominal
aorta at the beginning of the scan range with trigger-
ing threshold of 120 HU, or use manual triggering.
Contrast medium IV injection
Start delay
20 sec.*
Flow rate
3.5 ml/sec.
Total amount
120 ml
Abdomen
MIPthin: AngioFast
coronal
Image thickness
3
Distance between images
3
Number of images
35
For the 1
st
reconstruction the Autoload into MIPthin
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MIPthin, and a coronal
MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.