166
Pelvis
Body kernels:
• As standard kernels for body tissue studies B30s or
B40s are recommended; softer images are obtained
with B20s or B10s (extremely soft). The kernels B31s
or B41s have about the same visual sharpness as B30s,
respectively, B40s, the image appearance, however,
is more agreeable due to a ”fine-grained” noise struc-
ture; quite often, the low contrast detectability is
improved by using B31s, B41s instead of B30s, B40s.
• The 2 kernels B35s and B46s are especially designed
for cardiac application, in particular, B35s should be
used for Ca-scoring, and B46s for patency of stents.
• For higher sharpness, as is required e.g. in patient
protocols for cervical spine, shoulder, extremities,
thorax, lung, the kernels B50s, B60s, B70s, B80s are
available.
• A special high resolution lung kernel is supplied
with B90s; by using kernel B90s, practically the same
image impression is obtained as with HCE-filtering of
a B40s image.
Summary of Contents for SOMATOM Emotion 6
Page 1: ...SOMATOM Emotion 6 Application Guide Routine Protocols Software Version A70 ...
Page 23: ...23 General ...
Page 43: ...43 Children ...
Page 59: ...59 Children ...
Page 89: ...Head 89 ...
Page 102: ...102 Shoulder Overview Shoulder Spiral mode for HiRes bone studies and soft tissues ...
Page 139: ...139 Abdomen ...
Page 167: ...167 Pelvis ...
Page 175: ...175 Spine ...
Page 185: ...185 Extremities ...