147
AbdCombi
2
nd
reconstr.
kV
130
Effective mAs
130
Slice collimation
1.0 mm
Slice width
5.0 mm
1.25 mm
Feed/Rotation
9.0 mm
Rotation time
0.6 sec.
Kernel
B41s
B41s
Increment
5.0 mm
0.8 mm
Image order
cr-ca
CTDI
Vol
17.2 mGy
If you want to use this protocol for a two-phase study,
repeat the same protocol as mentioned below, and
chose start delay time for arterial phase as 20-25 s. In
this case, the thin slice reconstruction can also be used
for postprocessing. But do not administer oral contrast
medium, as this impairs the editing of MIP/SSD/VRT
images. Water could be used instead if necessary.
You can repeat the same protocol simply by clicking
the chronicle with the right mouse button for “repeat“.
E.g. when both non-contrast and contrast studies are
required.
Contrast medium IV injection
Start delay
50 – 65 sec.
Flow rate
4 ml/sec.
Total amount
100 ml
Abdomen
MIPthin: AbdCombi
coronal
Image thickness
3
Distance between images
3
Number of images
50
For the 2
nd
reconstruction the Autoload into the MIPthin
Range on the 3D Card is activated. The images will
be automatically loaded into 3D, MIPthin, and coronal
MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
Summary of Contents for SOMATOM Emotion 6
Page 1: ...SOMATOM Emotion 6 Application Guide Routine Protocols Software Version A70 ...
Page 23: ...23 General ...
Page 43: ...43 Children ...
Page 59: ...59 Children ...
Page 89: ...Head 89 ...
Page 102: ...102 Shoulder Overview Shoulder Spiral mode for HiRes bone studies and soft tissues ...
Page 139: ...139 Abdomen ...
Page 167: ...167 Pelvis ...
Page 175: ...175 Spine ...
Page 185: ...185 Extremities ...